Edge Ultrasound System Podręcznik Użytkownika Strona 166

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 165
154
(a)This index is not required for this operating mode; value is <1.
(b)This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason
listed. (Reference Global Maximum Index Value line.)
Data are not applicable for this transducer/mode.
Table 31: Transducer Model:
P10x
Operating Mode:
CW Doppler
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value (a) (a) 2.1 2.0
Associated Acoustic
Parameter
p
r.3
(MPa) #
W
0
(mW) —# 40.72 30.00
min of
[W
.3
(z
1
),I
TA.3
(z
1
)]
(mW)
z
1
(cm)
z
bp
(cm)
z
sp
(cm) # 0.7
d
eq
(z
sp
) (cm) 0.36
f
c
(MHz) # # 4.00 4.00
Dim of A
aprt
X(cm) # 0.320 0.16
Y(cm)
—# 0.70.7
Other Information
PD (μsec) #
PRF (Hz) #
p
r
@PII
max
(MPa) #
d
eq
@Pll
max
(cm) 0.27
Focal Length FL
x
(cm) —# 0.92
FL
y
(cm) —# 5.0
I
PA.3
@MI
max
(W/cm
2
)
#
Operating
Control
Conditions
Control 1: Exam Type Card Card
Control 2: Depth Any Any
Control 3: Zone Zone 3 Zone 0
Przeglądanie stron 165
1 2 ... 161 162 163 164 165 166 167 168 169 170 171 ... 197 198

Komentarze do niniejszej Instrukcji

Brak uwag